NewLab 410 – Freezing Point

Freezing Point in measure in aviation fuels, aviation gasoline, aviation turbine fuels, engine coolants, antifreeze products, brake fluids, etc.
Solidification Point of Benzene.
Solidification Point of Industrial Organic Chemicals.
According to the methods, the sample is cooled down and stirred. The solid hydrocarbon crystals formation is detected by means of a light beam throught fiber optic reflected on a mirror. As soon as crystals are detected, the sample is warmed up until their complete disappearance.

Measuring Parameters

  • Temperatures: in °C / °F
  • Measuring range: +80°C … -90°C (410/2 SA)
  • Resolution: 0.01 °C
  • Accuracy: ± 0.1 °C
  • Repeatability / Reproducibility: as per standards methods

Descripción

This unit meets specifications of the following methods: ASTM D852 – ASTM D1177 – ASTM D1493 – ASTM D1655 – ASTM D2386
ASTM D5901 – ASTM D5972 – ASTM D7154
IP 16 – IP 435 – IP 528 – IP 529
ISO 3013

Features:

New LabLink software able to manage up to 6 analytical heads simultaneously (stand alone)
User friendly interface
All analytical parameters recorded
Customizable analysis parameters and methods
Results report
Printable graphs and results – any Windows® compatible printer can be used